Radiation & Laser Safety
Radiation safety for XRF & laser safety for LIBS
Handheld XRF uses X-rays, handheld LIBS uses a laser. Both are safe when used correctly – but require knowledge, organisation and clear rules.
Part 1 · Handheld XRF
Radiation protection for X-ray fluorescence
Handheld XRF (HHXRF) devices generate ionising X-rays for non-destructive material analysis – e.g. alloy identification, scrap sorting, PMI or coating analysis. Because X-rays are produced, they fall under the German Radiation Protection Act and Ordinance (StrlSchG, StrlSchV).
Licence requirement
Expert inspection
Radiation protection responsible party (SSV)
Radiation protection officer (SSB)
Device training & expertise
Safe operation
Expertise per module R2.2 (Germany)
What is R2.2?
A course module under German radiation protection guidelines for operating X-ray equipment – relevant e.g. for X-ray fluorescence analysis. Contents: physics of X-rays, biological effects, dose limits, legal requirements, practical protection and fault response.
Who needs it?
For people to be appointed radiation protection officers for handheld XRF devices. Practical device training is acquired additionally, e.g. at delivery by the manufacturer.
How to obtain?
Through an accredited course (usually two days) with exam plus required practical experience. It must be refreshed every 5 years.
Documents for official recognition of the SSB
The radiation protection officer may only act officially after approval by the authority. Usually required:
- –Criminal record certificate
- –Proof of professional qualification
- –Radiation expertise certificate
- –Proof of device training
- –Written appointment by the company
Recurring obligations
These measures keep the device permanently safe and legally compliant:
| Measure | Interval |
|---|---|
| Expert inspection of the device | every 5 years |
| Refresher of radiation expertise | every 5 years |
| Instruction of all users | annually |
| Documentation of inspections | continuously |
| Checks of the device condition | regularly |
Part 2 · Handheld LIBS
Laser safety for LIBS
Handheld LIBS spectrometers produce no ionising radiation and therefore do not fall under radiation protection law, but under laser safety regulations. As portable LIBS devices typically use pulsed class 3B lasers, the German ordinance on artificial optical radiation (OStrV) applies.
Laser safety officer (LSB)
Risk assessment
Instruction
Notification & PPE
Class 3B device – operated under “class 1 conditions”
A laser safety officer cannot legally change the manufacturer's classification under DIN EN 60825-1 (e.g. class 3B) – the device class remains permanently. However, SciAps devices such as the Z-900 series use an integrated sample detection system (contact sensors): the high-energy class 3B laser only fires when the measuring nose lies completely flat on a sample. If the device is lifted, the laser stops immediately (safety interlock).
Practical relief in operation
If the LSB documents in the risk assessment that no accessible laser radiation is present in normal operation (thanks to the contact sensors), many strict requirements are waived: usually no laser safety eyewear for operator and bystanders (provided the sample seals tightly) and no cordoned-off laser area or laser cabin.
What remains mandatory
Since the device legally remains a class 3B product, the organisational minimum requirements of the OStrV still apply: a laser safety officer appointed in writing, annual instruction of employees specifically for this device, and formal notification to the employers' liability insurance association and the competent occupational safety authority.